Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0265-2027Class II

Generic Name: MiniMed Go App Generic Name: MiniMed Go App Model/Catalog Number: MMT-8600 (iOS), MMT-8601...

Medtronic MiniMed, Inc. / initiated 2026-09-02 / Open, Classified

Recall record

Recall number
Z-0265-2027
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Medtronic MiniMed, Inc.
Product
Generic Name: MiniMed Go App Generic Name: MiniMed Go App Model/Catalog Number: MMT-8600 (iOS), MMT-8601 (Android) Software Version: Software version 1.2.1 Product Description: The MiniMed Go app for iOS or MiniMed Go app for Android is compatible with the following components: " Simplera sensor " Instinct sensor " InPen smart insulin pen " Apple Watch [Optional] Continuous glucose monitoring (CGM) is a technology that allows you to continuously view glucose values. The app uses an inserted Simplera sensor CGM or Instinct sensor CGM with Bluetooth wireless technology that converts small amounts of glucose from the interstitial fluid under the skin into an electronic signal. The sensors use that signal to provide sensor glucose (SG) values to the MiniMed Go app on a compatible mobile device. The app also provides alerts based on your SG levels and contains a user interface for entering data such as insulin, meals, exercise, blood glucose (BG) values, and uploading information to the CareLink" Personal website. The app is available to download from the Apple"* App Store"* or the Google Play"* Store. Component: No
Product code
NDC Calculator, Drug Dose
Reason for recall
A software defect in which existing users who had previously configured and enabled the long-acting dose reminder feature in version 1.2.0 or earlier experienced failure of the long-acting dose reminder after updating their app to version 1.2.1. The long-acting dose reminder is an optional feature that, when enabled, triggers a notification at the scheduled time to remind users to take their long-acting insulin dose. The issue was caused by a type cast exception during database migration that occurs upon app update to version 1.2.1 causing a silent failure that prevented the creation of new alerts. Failure to receive the long-acting dose reminder may result in missed or delayed insulin doses, potentially leading to hyperglycemia. Conversely, if users are uncertain whether they took their dose due to the missing reminder, inadvertent duplicate dosing may occur, potentially resulting in hypoglycemia.
FDA root cause
Software design
Action
On 09/14/2026, customers were contacted via SMS, emailed or via an "URGENT MEDICAL DEVICE CORRECTION" Letter informing them that, an issue was identified with the MiniMed Go app in which the long-acting dose reminders were not being delivered after the app was updated to version 1.2.1. Actions required: Users of MiniMed Go app software versions 1.2.1 should update to version 1.2.2 or higher as soon as possible to resolve this issue. This update will ensure enabled long-acting reminders are delivered as expected. For iOS Users: To confirm the latest app version on your iOS device, follow the steps below: 1. Open the App Store. 2. Tap your profile icon at the top of the screen. 3. Scroll to see pending updates. 4. Tap Update next to the MiniMed Go app icon. For Android Users: To confirm the latest app version on your Android device, follow the steps below: 1. Open the Google Play Store. 2. Tap Menu, then select Manage apps & device. 3. Apps with an update available are labeled Update available. You can also search for theMiniMed Go app. 4. Tap Update next to the MiniMed Go app icon. For Questions or assistance - call the Medtronic MiniMed 24-Hour Technical Support line at 1-800-646-4633, option 1.
Quantity in commerce
623
Distribution
Worldwide - U.S. Nationwide distribution in the states of AL, CA, CT, FL, LA, MA, MD, NY, OH, PA, and TX. The countries of Austria, Belgium, Denmark, Finland, Germany, Iceland, Italy, Norway, Spain, Sweden, Switzerland, United Kingdom.
Product codes and lots
Lot Code: GTIN for MMT-8600, iOS: 00199150038978 GTIN for MMT-8601, Android: 00199150038985 UDI for MMT-8600, iOS: 00199150038978 UDI for MMT-8601, Android: 00199150038985 Software version 1.2.1

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0265-2027
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0265-2027Class IIOngoingVoluntary: Firm initiated