Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0284-2027

ETHILON Nylon Suture, Product Code 697G

Ethicon, LLC / initiated 2026-09-23 / Open, Classified

Recall record

Recall number
Z-0284-2027
Status
Open, Classified
Initiated
Posted
Recalling firm
Ethicon, LLC
Product
ETHILON Nylon Suture, Product Code 697G
Product code
GAQ Suture, Nonabsorbable, Steel, Monofilament And Multifilament, Sterile
Reason for recall
Nylon suture indicated for use in general soft tissue approximation and/or ligation may have needles with distorted curvature, that may break during use, or could cause additional soft tissue or vessel injury, including tearing or bleeding intraoperatively.
FDA root cause
Process control
Action
On 9/3/2026, field safety notices were mailed to customers who were asked to do the following: 1) Examine your inventory immediately to determine if you have product subject to this recall on hand and quarantine such product(s). If you have product subject to this recall, please maintain a copy of this notice with the quarantined product and keep a copy for your records. 2. Return affected product to firm. 3. Communicate the issue to relevant operating room or materials management personnel, or anyone else in your facility who needs to be informed. If any product subject to this recall has been forwarded to another facility, contact that facility to arrange return and forward them this recall notice. 4. Complete and return the Business Reply Form (BRF) via email to [email protected]
Quantity in commerce
4,003
Distribution
International distribution to the countries of JP, CA, AU.
Product codes and lots
GTIN Box/Unit: 30705031008568/10705031008564. Lot: 10B3QS