Z-0323-2027Class II
Illumina DRAGEN Server for NextSeq 550Dx REF: 20086130
Illumina, Inc. / initiated 2024-05-08 / Open, Classified
Recall record
- Recall number
- Z-0323-2027
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Illumina, Inc.
- Product
- Illumina DRAGEN Server for NextSeq 550Dx REF: 20086130,
- Product code
- PFF High Throughput Dna Sequence Analyzer
- Reason for recall
- (Retrospectively reported) Cybersecurity vulnerability identified in communication protocol between the customer-provided infrastructure (network storage, servers, or computers) and in vitro diagnostic devices which may result in compromised data transmission.
- FDA root cause
- Labeling design
- Action
- On May 8, 2024, Illumina, Inc. issued a Urgent Medical Device Recall via email. Illumina ask consignees to take the following 1. Download & read this Cybersecurity Guidance document. The full hyperlink is below. https://support.illumina.com/support-content/cyberguidance.html 2. For Each Product Connected to the Customer Network: work with your internal IT department to determine which secure protocol works best in your environment and to implement the appropriate configuration change. Note: a change to the network protocol affects both Dx and RUO modes simultaneously therefore there is no need to boot into each mode separately to take the necessary action. 3. Complete and return the Verification Form after carrying out all the steps in the instructions provided on your specific product(s) identified as affected in Table 1.
- Quantity in commerce
- 32 units
- Distribution
- Worldwide - US Nationwide distribution including in the states and territory of AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NH, NJ, NM, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV and the countries of AE, AR, AT, AU, BE, BG, CA, CH, CL, CN, CY, CZ, DE, DK, DZ, EG, ES, FI, FR, GB, Ghana, GR, GT, HK, HU, IE, IL, IS, IT, JP, KR, KW, LT, LU, LV, MA, MT, NG, NL, NO, NZ, PE, PH, PL, PT, QA, RO, RU, SA, SE, SG, SI, TH, TR, UA, VN, ZA.
- Product codes and lots
- All Lots/UDI: N/A
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0320-2027 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0321-2027 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0322-2027 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0323-2027 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0324-2027 | Class II | Ongoing | Voluntary: Firm initiated |