Z-0331-2027

Halyard MINOR PACK. Model/Catalog Number: VASF105-08

AVID Medical, Inc. / initiated 2026-09-08 / Open, Classified

Recall record

Recall number
Z-0331-2027
Status
Open, Classified
Initiated
Posted
Recalling firm
AVID Medical, Inc.
Product
Halyard MINOR PACK. Model/Catalog Number: VASF105-08.
Product code
LRO General Surgery Tray
Reason for recall
Potential for open seals on header bags, compromising sterility.
FDA root cause
Process control
Action
AVID Medical notified consignees on about 09/08/2026 via emailed "URGENT: MEDICAL DEVICE CORRECTION" letter. Consignees were instructed to Segregate, Quarantine, and Inspect all kits identified. If the header bag seal appears to be compromised, discard the entire kit. Consignees were also requested to complete and return the provided Recall Response Form. Distributors were instructed to notify customers if affected kits were further distributed. On 10/06/2026 AVID Medical sent an updated "URGENT: MEDICAL DEVICE CORRECTION" letter to consignees via email. The strategy was updated and consignees were instructed not to use the affected kits and that they should either discard or return them. If returning, they should contact customer service for a Return Goods Authorization. Consignees were again requested to complete and return the provided Recall Response Form. Distributors were again instructed to notify customers if affected kits were further distributed.
Quantity in commerce
108 kits
Distribution
U.S Nationwide distributor in the state of Illinois.
Product codes and lots
Model/Catalog Number: VASF105-08. UDI-DI: 10809160451960. Lot Number: 1659384. Expiration Date: 9/30/2027.