Z-0282-2027Class II
Halyard ANTERIOR HIP PACK. Model/Catalog Number: SOAH31
AVID Medical, Inc. / initiated 2026-09-08 / Open, Classified
Recall record
- Recall number
- Z-0282-2027
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- AVID Medical, Inc.
- Product
- Halyard ANTERIOR HIP PACK. Model/Catalog Number: SOAH31.
- Product code
- LRO General Surgery Tray
- Reason for recall
- A non-sterile component was piggybacked and incorrectly placed outside of the sterile barrier and not subjected to EO sterilization per kit specifications.
- FDA root cause
- Process control
- Action
- AVID Medical notified consignees on about 09/08/2026 via emailed "URGENT: MEDICAL DEVICE CORRECTION" letter. Consignees were instructed to identify, segregate, and quarantine all impacted kits identified, and to not use the impacted component located in the piggybacked portion of the convenience kit. Customers were also instructed to add warning labels to affected kits to ensure all users are aware of the actions to remove and discard the impacted component prior to use. distributors were instructed to notify consignees if affected units were further distributed. Additionally, consignees were requested to complete and return the Recall Response Form.
- Quantity in commerce
- 56 kits
- Distribution
- US Nationwide distributor in the state of Arizona.
- Product codes and lots
- Model/Catalog Number: SOAH31. UDI-DI: 10809160512890. Lot Code: Lot Number: 1689215. Expiration Date: 11/06/2026.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0282-2027 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0283-2027 | Class II | Ongoing | Voluntary: Firm initiated |