Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0283-2027Class II

Halyard ANTERIOR HIP PACK. Model/Catalog Number: SOAH30

AVID Medical, Inc. / initiated 2026-09-08 / Open, Classified

Recall record

Recall number
Z-0283-2027
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
AVID Medical, Inc.
Product
Halyard ANTERIOR HIP PACK. Model/Catalog Number: SOAH30.
Product code
LRO General Surgery Tray
Reason for recall
A non-sterile component was piggybacked and incorrectly placed outside of the sterile barrier and not subjected to EO sterilization per kit specifications.
FDA root cause
Process control
Action
AVID Medical notified consignees on about 09/08/2026 via emailed "URGENT: MEDICAL DEVICE CORRECTION" letter. Consignees were instructed to identify, segregate, and quarantine all impacted kits identified, and to not use the impacted component located in the piggybacked portion of the convenience kit. Customers were also instructed to add warning labels to affected kits to ensure all users are aware of the actions to remove and discard the impacted component prior to use. distributors were instructed to notify consignees if affected units were further distributed. Additionally, consignees were requested to complete and return the Recall Response Form.
Quantity in commerce
80 kits
Distribution
US Nationwide distributor in the state of Arizona.
Product codes and lots
Model/Catalog Number: SOAH30. UDI-DI: 10809160520246. Lot Code: Lot Numbers: 1705826, 1705827. Expiration Date: 1/31/2028.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0283-2027
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0282-2027Class IIOngoingVoluntary: Firm initiated
Z-0283-2027Class IIOngoingVoluntary: Firm initiated