PENTAX Medical Ultrasound Endoscope. Model Numbers. EG38-J10UT, EG34-J10U

Pentax of America Inc / initiated 2026-09-09 / Open, Classified

Recall record

Recall number
Z-0307-2027
Status
Open, Classified
Initiated
Posted
Recalling firm
Pentax of America Inc
Product
PENTAX Medical Ultrasound Endoscope. Model Numbers. EG38-J10UT, EG34-J10U.
Product code
ODG Endoscopic Ultrasound System, Gastroenterology-Urology
Reason for recall
Warnings added to the IFU for ultrasound endoscopes and water bottle assembly to ensure there is deaerated sterile water in the assembly prior to and during the procedure.
FDA root cause
Unknown/Undetermined by firm
Action
An Urgent Field Safety Notice dated 09/02/2026 was mailed to consignees. The notice informs consignees of two updates to the IFU of PENTAX Medical products, specifically the ultrasound endoscope devices and water bottle assembly. Consignees are to inform their staff of this notice and include the warning and updates to their IFUs. Additionally, consignees are instructed to change water for each procedure and fill the bottle approximately 2/3 full, or to the 200 marking on the bottle scale, with sterile water. If the water/balloon feeding function does not work consignees are to fill the water bottle with deaerated sterile water. PENTAX Medical can be contacted by phone at 1-800-431-5880 Monday through Friday from 8:30 AM EST to 5:00 PM EST, or by email at [email protected].
Quantity in commerce
182 units
Distribution
Worldwide - US Nationwide and International distribution - Pending.
Product codes and lots
1. Model No. EG38-J10UT; UDI-DI: 04961333239474, 04961333239481, 04961333239498, 04961333239504, 04961333239528, 04961333252770; All Serial Numbers. 2. Model No. EG34-J10U; UDI-DI: 04961333239672, 04961333239689, 04961333239696, 04961333239702, 04961333239726, 04961333248834, 04961333248841, 04961333248858, 04961333248872, 04961333248896, 04961333252732; All Serial Numbers.