Z-0286-2027

Orthoscan VERSA Accessory Cart, REF: 1000-0021 and Versa Mini C-Arm System, REF: 1000-0020

Ziehm-Orthoscan Inc / initiated 2026-09-03 / Open, Classified

Recall record

Recall number
Z-0286-2027
Status
Open, Classified
Initiated
Posted
Recalling firm
Ziehm-Orthoscan Inc
Product
Orthoscan VERSA Accessory Cart, REF: 1000-0021 and Versa Mini C-Arm System, REF: 1000-0020
Product code
OXO Image-Intensified Fluoroscopic X-Ray System, Mobile
Reason for recall
The accessory cart weldment does not conform to weld dimensional specifications which may result in separation of the monitor arm mounting structure from the cart.
FDA root cause
Process control
Action
On September 11, 2026, Ziehm-Orthoscan issued a Urgent: Medical Device Recall Notification via Email. Ziehm-Orthoscan ask consignees to take the following actions: 1. The device may continue to be used until the corrective action has been completed, provided the user follows the precautions described in this notification and no signs of instability or unintended movement are observed. 2. Users should exercise caution when positioning, moving, or operating the system and monitor assembly, avoid fully extending the display arm until the corrective action has been performed. 3. If unintended movement, instability, loosening, or any other condition related to the issue described in this notification is observed, discontinue use of the affected system and contact Ziehm- Orthoscan immediately for further instructions. 4. Please ensure that a copy of this notification is made available to all users and operators of the affected device. It is recommended that this notification be posted in a visible location near the device until the corrective action has been completed. 5. If the affected device has been transferred, sold, or relocated to another facility or department, please ensure that a copy of this notification is provided to the current user or owner of the device. 6. Please maintain awareness of this notification until the corrective action has been completed and verified by Ziehm-Orthoscan or its authorized representative.
Quantity in commerce
55 units
Distribution
Worldwide - US Nationwide distribution in the states of AZ , CA , CO , CT , FL , GA , IL , IN , KS, MA , MD , MI , MN , NC , NE , NV , NY , OH , OR , SC, TX , UT , WA and the countries of Germany , Australia , Spain , Switzerland, Japan.
Product codes and lots
UDI/ Mini C-Arm serial number (6Axxxx)/Accessory cart serial Number: (VCXXX): 1 6A0110 /VC0110 2 6A0111 /VC0111 3 6A0112 /VC0112 4 6A0115 /VC0113 5 6A0114 /VC0114 6 6A0117 /VC0115 7 6A0116 /VC0116 8 6A0113 /VC0117 9 6A0118 /VC0118 10 6A0119/ VC0119, VC0120; 11 6A0120 /VC0121; 12 6A0121 /VC0122; 13 6A0123 /VC0123; 14 6A0122 /VC0124; 15 6A0126 /VC0125; 16 6A0133 /VC0134; 17 6A0124 /VC0126; 18 6A0128 /VC0128; 19 6A0127 /VC0129; 20 6A0129 /VC0130; 21 6A0125 /VC0127; 22 6A0131 /VC0132; 23 6A0134 /VC0133; 24 6A0142 /VC0144; 25 6A0136 /VC0136; 26 6A0132 /VC0137; 27 6A0137 /VC0138; 28 6A0138 /VC0139; 29 6A0139 /VC0140; 30 6A0140 /VC0141; 31 6A0141 /VC0142; 32 6A0146 /VC0147; 33 6A0143 /VC0143; 34 6A0144 /VC0145; 35 6A0152 /VC0153; 36 6A0147 /VC0148; 37 6A0148 /VC0149; 38 6A0149 /VC0150; 39 6A0150 /VC0151; 40 6A0151 / VC0152; 41 6A0163 / VC0165; 42 6A0130 / VC0131; 43 6A0155 / VC0156; 44 6A0157 / VC0157; 45 6A0156/ VC0158; 46 6A0158 / VC0159, VC0160; 47 6A0159/ VC0161; 48 6A0160/ VC0162; 49 6A0161 / VC0163; 50 6A0162 VC0164; 51 6A0153 VC0154; 52 6A0164 VC0166; 53 6A0165 VC0167