CryoPatch SynerGraft (SG) Pulmonary Human Cardiac Patch, Model: SGP010

Artivion, Inc / initiated 2026-09-18 / Open, Classified

Recall record

Recall number
Z-0325-2027
Status
Open, Classified
Initiated
Posted
Recalling firm
Artivion, Inc
Product
CryoPatch SynerGraft (SG) Pulmonary Human Cardiac Patch, Model: SGP010
Product code
DXZ Patch, Pledget And Intracardiac, Petp, Ptfe, Polypropylene
Reason for recall
A cryopreserved pulmonary human cardiac patch, taken from a 64-year-old human donor, was outside the cardiac tissue donor age specification of newborn to 60 years. The patch was implanted, and due to limitations of long-term in vivo data, the possibility of premature graft dysfunction cannot be excluded.
FDA root cause
Nonconforming Material/Component
Action
On 9/18/2026, the notification was delivered via mail to customer who was advised of the following: Available biomechanical and histological evidence does not indicate an increased risk that would warrant additional patient testing, enhanced clinical monitoring, or other clinical intervention beyond the patient's routine standard of care. Accordingly, prophylactic removal or surgical replacement of the already implanted tissue solely because of the donor-age nonconformance is not clinically warranted. Complete and return the response form via email to [email protected] Firm requests that any observed deficiency, performance concern, or other issue potentially associated with the tissue be reported to them vial email: [email protected] or phone: +1-770-419-3355
Quantity in commerce
1
Distribution
US Nationwide distribution in the state of IL.
Product codes and lots
Serial/Donor Number: 12617554/177871, UDI-DI: 00877234000478, Shipped: 22May2024, Implanted: 05June2024