Z-0332-2027Class II
TruSystem 7000 U Operating Table (dV), Product Code 1723633
Baxter Healthcare Corporation / initiated 2026-09-14 / Open, Classified
Recall record
- Recall number
- Z-0332-2027
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Baxter Healthcare Corporation
- Product
- TruSystem 7000 U Operating Table (dV), Product Code 1723633
- Product code
- GDC Table, Operating-Room, Electrical
- Reason for recall
- The main lift may unexpectedly drop from its set position to its lowest position during use. The tabletop may not be able to be raised again after the drop occurs. An uncontrolled drop of the surgical table may result in an abrupt patient displacement during surgery. Serious or critical adverse health consequences may occur, including traumatic injury, airway compromise, and surgical procedure complications, among others.
- FDA root cause
- Equipment maintenance
- Action
- Customers were notified by telephone beginning 9/14/26. Baxter Healthcare Corporation is issuing a voluntary correction for the TruSystem Surgical Tables after identifying that four units underwent spindle replacement without the required belt tension verification. The affected units were released without confirmation that belt tension met the specified requirements, which could lead to unintended downward movement of the table column. Unintended downward movement of the surgical table could occur, potentially resulting in disruption of patient positioning, reduced perfusion or oxygenation, musculoskeletal injury and/or interruption of a surgical procedure. To date, Baxter has not received any reports of patient injury related to this issue. Baxter will call to schedule a time to come in and perform the belt tension verification. Customers can continue to use the tables.
- Quantity in commerce
- 62 units
- Distribution
- US Nationwide distribution in the states of IL, MO, TX.
- Product codes and lots
- UDI-DI: 00887761967519; Serial Number: 110188921, 110188922, and 110206662
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0332-2027 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0333-2027 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.