Z-0341-2027
Generic Name: Perma Sharp Sutures; Generic Name: Polypropylene Nonabsorbable Surgical Suture; Model/Catalog...
Henry Schein Inc. / initiated 2026-09-17 / Open, Classified
Recall record
- Recall number
- Z-0341-2027
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Henry Schein Inc.
- Product
- Generic Name: Perma Sharp Sutures; Generic Name: Polypropylene Nonabsorbable Surgical Suture; Model/Catalog Number: PSN8660P (HSI Item No. 600-7193); Product Description: Sterile, synthetic, nonabsorbable, monofilament 6-0 blue polypropylene surgical suture, 18-inch length, with C-22 16.2 mm, 3/8-circle reverse-cutting needle. Component: N/A;
- Product code
- GAW Suture, Nonabsorbable, Synthetic, Polypropylene
- Reason for recall
- Products lack 510(k) clearance.
- FDA root cause
- Process control
- Action
- On September 25, 2026, URGENT: MEDICAL DEVICE RECALL NOTIFICATION letters were sent to customers. ACTION: Please examine your inventory to determine whether you have affected product. If affected product is identified, immediately cease distribution and use, remove the product from available inventory, and follow the return instructions provided. The product must be returned within 30 days to Henry Schein s Returns Facility located at: 41 Weaver Rd, Denver, PA 17517.
- Quantity in commerce
- 73
- Distribution
- Worldwide distribution - U.S. Nationwide and the countries of Canada, Uganda, and the Cayman Islands.
- Product codes and lots
- Model/Catalog Number: PSN8660P (HSI Item No. 600-7193); UDI-DI: 10889950001896; Lot Numbers: 0624002HF, 0625007HF, 0625017HF, 0625020HF, 0625025HF, 0625026HF;