Z-0347-2027Class II
Trinias Opera Smart (C16s, F12s)/Opera Grand(C12s,C16s, F12s)/Smart OBL Interventional Fluoroscopic X-Ray...
Shimadzu Corporation Medical Systems Division / initiated 2025-09-30 / Open, Classified
Recall record
- Recall number
- Z-0347-2027
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Shimadzu Corporation Medical Systems Division
- Product
- Trinias Opera Smart (C16s, F12s)/Opera Grand(C12s,C16s, F12s)/Smart OBL Interventional Fluoroscopic X-Ray System containing C-arm support (MH-500, MH-60),
- Product code
- OWB Interventional Fluoroscopic X-Ray System
- Reason for recall
- The mounting bolts securing the C-arm drive unit of the X-ray system may not be torqued to specification, potentially compromising structural stability and safe operation.
- FDA root cause
- Process control
- Action
- On September 30, 2026 Shimadzu Corp. issued a Urgent Medical Device Recall Notification to affected consignees via email. Shimadzu asked consignees to take the following actions; 1. If the bolts become loose, an abnormal noise will occur when moving the C-arm. Please check for abnormal noises during pre-study inspections. 2. If an abnormal noise is heard, please stop using the product and contact our service department. 3. Please complete the Medical Device Recall Customer Acknowledgement.
- Quantity in commerce
- 118 unts.
- Distribution
- US Nationwide distribution in the states of SC, TX, FL, TN, CA, FL, NY.
- Product codes and lots
- Serial number/Shimadzu Medical Systems #/UDI: M1F0D3BF4001/ SMS20250167/ (01)04540217067145(11)250; M1F0CC4EB001/ SMS20250098/ (01)04540217067022(11)241; M1F0D27EA001/ SMS20250001/ (01)04540217067138(11)241; 41F0F43E7001/ SMS20250034/ (01)04540217068951(11)240; M1F0D3BE7001/ SMS20250248/ (01)04540217067145(11)240; M1F0D3BE6001/ SMS20240196/ (01)04540217067145(11)240; 41F0F40E5001/ N/A/ (01)04540217068937(11)240; 41F0F41E5001/ N/A/ (01)04540217068944(11)240; 41F0F43E5001/ SMS20240236/ (01)04540217068951(11)240; M1F0CD7E4001/ SMS20240162/ (01)04540217067046(11)240; 41F0F41E1001/ SMS20240062/ (01)04540217068944(11)240; 41F0F43DB002/ SMS20240101/ (01)04540217068951(11)231; 41F0F43DB001/ SMS20240100/ (01)04540217068951(11)231; M1F0CD7D4001/ SMS20230178/ (01)04540217067046(11)230
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0347-2027 | Class II | Ongoing | Voluntary: Firm initiated |