Z-0287-2027Class I
Generic Name: Edwards Commander M Delivery System Generic Name: Edwards Commander M Delivery System...
Edwards Lifesciences, LLC / initiated 2026-10-05 / Open, Classified
Recall record
- Recall number
- Z-0287-2027
- Classification
- Class I
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Edwards Lifesciences, LLC
- Product
- Generic Name: Edwards Commander M Delivery System Generic Name: Edwards Commander M Delivery System Model/Catalog Number: 9880CM29A Software Version: Not applicable. Product Description: Prosthesis, mitral valve, percutaneously delivered Component: No
- Product code
- NPU Prosthesis, Mitral Valve, Percutaneously Delivered
- Reason for recall
- Due to a manufacturing issue which may result in cardiovascular balloon rupturing.
- FDA root cause
- Process change control
- Action
- On 10/05/2026, an "URGENT: MEDICAL DEVICE RECALL (PRODUCT REMOVAL)" Letter was sent via various delivery methods (e.g., electronic delivery (e-mail), FedEx Priority Overnight mail, or on-site visit) to customers informing them that, due to complaints certain manufacturing lots of the SAPIEN M3 Commander M delivery systems potentially contain a manufacturing defect, which may result in balloon burst during use. Customers are instructed to: " Review this notice carefully and share it with the appropriate clinical staff at their site. " Immediately examine their inventory and quarantine units subject to this recall. Do not use these affected units. Units with other lot numbers are not impacted by this recall and can continue to be used. " Complete the attached Customer Acknowledgment Form and return it to Edwards TMTT Representative or via email to [email protected] " Contact Customer Service at 1-800-424-3278 or email: [email protected] to obtain a Returned Goods Authorization (RGA) number and return any impacted product to Edwards. Edwards asks that customers complete the last two actions listed above within 15 days of receipt of this notification. Customers' assistance is crucial to ensuring this recall is executed thoroughly and promptly. This notice has been communicated by Edwards to the appropriate Regulatory Authorities. For questions and further assistance, contact an Edwards representative who can assist with product return.
- Quantity in commerce
- 164
- Distribution
- U.S.: AZ, CA, CO, FL, GA, IL, KS, LA, MA, MI, MN, MO, NC, NJ, NY, OH, OK, PA, SC, TN, TX, UT, VA and WA. O.U.S.: N/A
- Product codes and lots
- Lot Code: Model No.: 9880CM29A; UDI-DI: 00690103219777; Recalled Lot Number(s): 67718561, 67804014, 67887043, 67887044.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0287-2027 | Class I | Ongoing | Voluntary: Firm initiated |