Z-0349-2027Class II

Cochlear Nucleus Nexa cochlear implant with Contour Advance electrode REF P1979576 Model Number: CI1000...

Cochlear Limited / initiated 2026-09-30 / Open, Classified

Recall record

Recall number
Z-0349-2027
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Cochlear Limited
Product
Cochlear Nucleus Nexa cochlear implant with Contour Advance electrode REF P1979576 Model Number: CI1000 UDI-DI codes: 9321502069847 Cochlear Nucleus CI1000 Series implants are prescription only, single use devices intended for long term implantation under the skin in the mastoid region on either side of the head. The cochlear implant system works by converting sound in the environment into electrical code. It transmits this code to the auditory nerve and on to the brain where it is interpreted as sound. The Cochlear Implant System consists of internal (implanted) and external components. The implanted component consists of a receiver/stimulator and an electrode array. The external components consist of a sound processor and their associated transmitting coils and cables and must be worn for the implant to work.
Product code
MCM Implant, Cochlear
Reason for recall
Due to a manufacturing deviation of incomplete electrode welds of cochlear implants and resulting in one or more open circuits
FDA root cause
Under Investigation by firm
Action
On 09/30/2026, an "URGENT Field Safety Corrective Action Notification" was sent to U.S. customers informing them that a potential condition has been identified that could result in one or more open circuits affecting individual electrode pathways with the implant. Customers were instucted to not implant and return affected products to Cochlear Limited. For questions contact [email protected].
Quantity in commerce
4 device in the U.S.
Distribution
U.S.: NJ, NY, and PA. O.U.S.: Not provided
Product codes and lots
Model: CI1000 Product/Catalog: P1979576 UDI-DI code: 9321502069847 Serial Numbers: 1020210051972 1020210070461 1020210145046 1020210048286

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0349-2027
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0349-2027Class IIOngoingVoluntary: Firm initiated