Z-0355-2027
Yeastone Broth, 11ML, 10/BOX, used in antibiotic susceptibility testing of mycological clinical specimens
Remel, Inc / initiated 2026-10-19 / Open, Classified
Recall record
- Recall number
- Z-0355-2027
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Remel, Inc
- Product
- Yeastone Broth, 11ML, 10/BOX, used in antibiotic susceptibility testing of mycological clinical specimens
- Product code
- NGZ Susceptibility Test Plate, Antifungal
- Reason for recall
- Affected lot did not meet specifications. Product may report false susceptibility to antifungal agents, specifically within the Azole family.
- FDA root cause
- Under Investigation by firm
- Action
- The firm began notifying affected customers on October 19, 2026. Customers were instructed to notify any personnel who needed to be aware of the issue. Customers should review previously reported susceptibility results generated using the affected lot and assess whether retesting is appropriate. Customers should destroy any remaining inventory of the affected lot. Do not use it for any clinical laboratory testing. Customers should complete and return the acknowledgement form accompanying the letter. Firm will replace product or issue credit. If you have any questions, please contact our Technical Services Department at 800-255-6730 (US) or email [email protected]. You should complete the accompanying Acknowledgment Form upon receipt and return to [email protected]
- Quantity in commerce
- 33 US, 54 OUS
- Distribution
- Worldwide - US Nationwide distribution in the states of MA and CA. The countries of India, Saudi Arabia, UK.
- Product codes and lots
- YY3462 lot 440400 UDI-DI 848838018817