Senographe Pristina 3D

GE Medical Systems, LLC / initiated 2026-08-28 / Open, Classified

Recall record

Recall number
Z-0328-2027
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
GE Medical Systems, LLC
Product
Senographe Pristina 3D
Product code
OTE Digital Breast Tomosynthesis
Reason for recall
GE HealthCare has become aware of a nonconformance to certain Senographe Pristina VIA systems. Specifically, in cases of 3D+2D manual mode where the acquisition configuration of the anode track/filter combination is different, it can result in the following: the collimator blades might be visible in the image, or the collimator might be open more than expected, and an incorrect collimator filter might be used with respect to the anode track selected This issue could result in a minor additional X-ray dose being delivered to the patient and obvious artifacts in the image (collimator blades).
FDA root cause
Radiation Control for Health and Safety Act
Action
GE HealthCare will send a customer letter dated August 28, 2026, that provides safety instructions for immediate mitigation of the incorrect configuration of the collimator anode track / filter in manual mode during 3D+2D acquisitions. GE HealthCare will correct all affected Senographe Pristina VIA systems by upgrading the software at no cost to the customer. If you have any questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.
Distribution
USA and OUS.
Product codes and lots
N/A

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0328-2027
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0326-2027Class IIOngoingFDA Mandated
Z-0327-2027Class IIOngoingFDA Mandated
Z-0328-2027Class IIOngoingFDA Mandated