Z-0421-03

AxSYM System, Lift Mechanisms: List Numbers 7A83-01, 7A83-95, 7A83-98, and 7A83-03

Abbott Medical / initiated 2002-12-18 / Terminated

Abbott Laboratories, Inc began a recall of AxSYM System, Lift Mechanisms: List Numbers 7A83-01, 7A83-95, 7A83-98, and 7A83-03 on . The reason FDA records is "Gas Springs supporting processing center cover lose pressure after repeated usage causing the cover to unexpectedly close". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0421-03
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
AxSYM System, Lift Mechanisms: List Numbers 7A83-01, 7A83-95, 7A83-98, and 7A83-03
Product code
JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Reason for recall
Gas Springs supporting processing center cover lose pressure after repeated usage causing the cover to unexpectedly close.
FDA root cause
Other
Action
Abbott Laboratories sent a December 18, 2002 letter to all consignees informing consignees of their plans to install a locking mechanism on all AxSYM Systems. The installations will be performed by Abbott Representatives. A customer reply form was included in the device correction letter.
Quantity in commerce
18,832 units
Distribution
Nationwide and to the following foreign countries: Mexico, Guatemala, Venezuela, Columbia, Ecuador, Peru, Chile, Uruguay, Argentina, Brazil, Dominican Republic, El Salvador, Canada, Germany, Ireland, England, Belgium, France, Austria, Italy, Spain, Russia, Japan, Singapore, Hong Kong, Taiwan, Thailand, Malaysia, Korea, Australia, New Zeland, Honduras, Costa Rica, Trinidad, Bahamas, Northern Antilles, Bermuda, Panama, West Indies, Paraguay, Norway, Kingstow, VC
Product codes and lots
List Numbers: 7A83-01, 7A83-95, 7A83-98, and 7A83-03