Z-0420-03

Pharmaseal Thoracentesis Tray with Catheter, catalog number 4341B; a sterile, single use, Rx, latex free...

Medtronic, Inc. / initiated 2002-11-15 / Terminated / root cause: manufacturing process

Cardinal Health began a recall of Pharmaseal Thoracentesis Tray with Catheter, catalog number 4341B; a sterile, single use, Rx, latex free procedure tray containing a 7'' long Thoracentesis... on . The reason FDA records is "Raw material used in the manufacture of catheters may contain a contaminant predisposing the catheter to become brittle, possibly breaking during use". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0420-03
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic, Inc.
Product
Pharmaseal Thoracentesis Tray with Catheter, catalog number 4341B; a sterile, single use, Rx, latex free procedure tray containing a 7'' long Thoracentesis Catheter with a 14 gauge x 2'' needle, catheter guard, 60 ml syringe, needles, 5 ml syringe, drainage tube with needle, fluid collection bag, three prelabeled specimen vials with caps, 2 swab sticks, three gauze pads, fenestrated drape with adhesive strips, towel, 5 ml -1% lidocaine hydrochloride, hospital wrap and bandage; Allegiance Healthcare Corporation, McGaw Park, IL 60085 USA
Product code
GCB--
Reason for recall
Raw material used in the manufacture of catheters may contain a contaminant predisposing the catheter to become brittle, possibly breaking during use.
Root cause tag
manufacturing process
FDA root cause
Other
Action
Recall letters dated 11/15/02 were sent via 2-day FedEx to the direct accounts. The accounts were informed of the potential for breakage of the catheters during use, requested to inspect their inventories for the affected lot numbers, and segregate any found for return to Allegiance for replacement. Distributors were requested to sub-recall the product from their accounts. The recall was extended to include the 21 additional lots by letter dated 4/8/04. The letters were sent via Airborne Express, 1-Day, on 4/8/04. The instructions were the same as the 11/15/02 letter, except that Allegiance has changed their name to Cardinal Health.
Quantity in commerce
232,228 trays
Distribution
Nationwide distribution, including Puerto Rico and Guam, and internationally to Saudi Arabia
Product codes and lots
L1J035X, L1J088, L1K042, L1K058, L1K070, L1L057X, L1N018, L1N031, L1N039X, L1N094, L1P017, L1P045, L1S015, L2A012, L2A021, L2A048, L2A075, L2B039, L2B047 ,L2B073, L2C012, L2C038, L2C083, L2D041, L2D067, L2E036  The recall was expanded on 4/8/04 to include the following additional lots: L1J072, L2D102X, L2H001, L2H007, L2J004, L2J021, L2J088, L2K001, L2K021, L2L026, L2L037, L2N022, L2N085, L2P113, L2S123X, L2S128, L2S166, L2S185, L3A105, L3A218X, L3A224.