Baxter Meridian Hemodialysis Instrument, product code 5M5576 and 5M5576R; Baxter Healthcare Corporation...
Baxter Healthcare Corp. began a recall of Baxter Meridian Hemodialysis Instrument, product code 5M5576 and 5M5576R; Baxter Healthcare Corporation, McGaw Park, IL 60085 The instrument is part of a high... on . The reason FDA records is "Meridian hemodialysis instrument may reset when in shutdown mode causing the system to reset with potential for patient increased bioburden exposure". As of , the recall status on record is Terminated, and MDM tags the root cause as sterility.
Recall record
- Recall number
- Z-0605-03
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Baxter Healthcare Corporation
- Product
- Baxter Meridian Hemodialysis Instrument, product code 5M5576 and 5M5576R; Baxter Healthcare Corporation, McGaw Park, IL 60085 The instrument is part of a high permeability hemodialysis system which consists of a controlled dialysate delivery system that incorporates an ultrafiltration controller to prevent excessive loss of water from the patient''s blood, an extracorporeal blood set, and a high permeability dialyzer. The standard features of the Meridian instrument incluide a high blood flow rate capacity (for shortened hemodialysis treatment time), automatic ultrafiltration control, standard and variable bicarbonate and sodium capabilities anbd automated chemical disinfection. The Meridian will operate in either the bicarbonate or acetate mode for concentrates.
- Product code
- KDI Dialyzer, High Permeability With Or Without Sealed Dialysate System
- Reason for recall
- Meridian hemodialysis instrument may reset when in shutdown mode causing the system to reset with potential for patient increased bioburden exposure.
- Root cause tag
- sterility
- FDA root cause
- Other
- Action
- Baxter issued a Safety Alert letter dated 11/15/02 to the direct accounts informing them of the potential for incomplete disinfection of the Meridian instrument during the Renalin/Formalin disinfection cycle due to electomagnetic interference (EMI). The accounts were informed that they will be receiving a radio frequency interference (RFI) filter kit and installation instructions to eliminate EMI by 2/28/03. The accounts were given instructions to follow to ensure that the Renalin/Formalin disinfection cycle has been completed prior to use of the machine.
- Quantity in commerce
- 2864 units
- Distribution
- Nationwide, Mexico, China, South Korea, Hong Kong
- Product codes and lots
- product codes 5M5576, 5M5576R, serial numbers 200000 through 202864