Z-0521-03
Baxter 60'' Micro-Volume Extension Set, catalog 2N3350; Baxter Healthcare Corporation, Deerfield, IL 60015
Baxter Healthcare Corporation / initiated 2002-10-17 / Terminated / root cause: labeling and UDI
Baxter Healthcare Corp. began a recall of Baxter 60'' Micro-Volume Extension Set, catalog 2N3350; Baxter Healthcare Corporation, Deerfield, IL 60015 on . The reason FDA records is "The I.V. filter sets contain a 1.2 micron filter instead of the labeled 0.22 micron filter". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0521-03
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Baxter Healthcare Corporation
- Product
- Baxter 60'' Micro-Volume Extension Set, catalog 2N3350; Baxter Healthcare Corporation, Deerfield, IL 60015
- Product code
- FPA Set, Administration, Intravascular
- Reason for recall
- The I.V. filter sets contain a 1.2 micron filter instead of the labeled 0.22 micron filter.
- Root cause tag
- labeling and UDI
- FDA root cause
- Other
- Action
- Recalled by letter dated 10/17/02. Accounts were informed of the potential for the filters to be 1.2 micron filters instead of the 0.22. micron filters as labeled. The accounts were requested to examine their inventories for the affected catalog numbers and destroy the units found for credit.
- Quantity in commerce
- 630,140 units
- Distribution
- Nationwide, Australia, Puerto Rico, Brazil, Canada, Sweden, New Zealand, Japan, Mexico, American Samoa, El Salvador, Turkey, Saudi Arabia, Austria, Cayman Islands, Macedonia, Honduras, Kuwait, Hong Kong and Nigeria.
- Product codes and lots
- All lots received prior to 10/23/2002