Z-0341-03

Pulse Generators (aka pacemakers) Models include: Z-031-1 Meta DDR, Model 1256D Z-032-1 Tempo VR, Model 1102...

Abbott Medical / initiated 2002-11-04 / Terminated

St Jude Medical began a recall of Pulse Generators (aka pacemakers) Models include: Z-031-1 Meta DDR, Model 1256D Z-032-1 Tempo VR, Model 1102 Z-033-1 Tempo V, Model 1902 Z-034-1 Tempo DR... on . The reason FDA records is "Potential loss of pacing function. Expansion of recalls Z-031-1 thru Z-036-1". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0341-03
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
Pulse Generators (aka pacemakers) Models include: Z-031-1 Meta DDR, Model 1256D Z-032-1 Tempo VR, Model 1102 Z-033-1 Tempo V, Model 1902 Z-034-1 Tempo DR, Model 2101 Z-035-1 Tempo D, Model 2902 Z-036-1 Meta, Model 1256
Product code
DXY Implantable Pacemaker Pulse-Generator
Reason for recall
Potential loss of pacing function. Expansion of recalls Z-031-1 thru Z-036-1
FDA root cause
Other
Action
Firm issued 'Dear Doctor' letters to following physicians which listed patient''s that the Doctor was responsible for. The letter informs of the situation regarding the potential loss of pacing function and recommends that the professional healthcare provider review and evaluate patients with the affected devices as soon as possible. Additional conditions to consider are specified for pacemaker dependent versus nondependent patients.
Quantity in commerce
1146
Distribution
Nationwide and Australia, Austria, Belgium, Brazil, Canada, Cyprus, Czech Rep., France, Germany, Italy, Netherlands, Spain, Turkey, Venezuela.
Product codes and lots
Serial numbers U6227335 to U6240907