COULTER LH 700 Series Hematology Analyzers

Siemens Healthcare Diagnostics Inc. / initiated 2002-11-08 / Terminated / root cause: software

Beckman Coulter Inc began a recall of COULTER LH 700 Series Hematology Analyzers on . The reason FDA records is "Software". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0416-03
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
COULTER LH 700 Series Hematology Analyzers
Product code
GKZ Counter, Differential Cell
Reason for recall
Software.
Root cause tag
software
FDA root cause
Other
Action
Customers received a telephone call and a faxed Product Corrective Action letter dated Nov. 8, 2002 which includes modified operating instructions. Additionally an updated software version was released and installed on all affected customers units. This recall is complete.
Quantity in commerce
132
Distribution
Nationwide and Canada
Product codes and lots
Part number -6605632, No lot number applies, all serial numbers apply.