Z-0718-03
TransportAir, Accessory to an Air Bed, Auxiliary Blower Unit, manufactured by Kinetic Concepts, Inc
KCI USA, Inc. / initiated 1997-06-01 / Terminated
Kinetic Concepts, Inc began a recall of TransportAir, Accessory to an Air Bed, Auxiliary Blower Unit, manufactured by Kinetic Concepts, Inc on . The reason FDA records is "Battery out-gassing due to battery over charging and/or secondary failure mode of deep-discharge". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0718-03
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- KCI USA, Inc.
- Product
- TransportAir, Accessory to an Air Bed, Auxiliary Blower Unit, manufactured by Kinetic Concepts, Inc.
- Product code
- INX Bed, Air Fluidized
- Reason for recall
- Battery out-gassing due to battery over charging and/or secondary failure mode of deep-discharge.
- FDA root cause
- Other
- Action
- In approximately November 1997 the firm implemented the TransportAir retrograde for all active units.
- Quantity in commerce
- 4264
- Distribution
- Nationwide
- Product codes and lots
- Serial numbers 0550037 through 0559399.