Z-0718-03

TransportAir, Accessory to an Air Bed, Auxiliary Blower Unit, manufactured by Kinetic Concepts, Inc

KCI USA, Inc. / initiated 1997-06-01 / Terminated

Kinetic Concepts, Inc began a recall of TransportAir, Accessory to an Air Bed, Auxiliary Blower Unit, manufactured by Kinetic Concepts, Inc on . The reason FDA records is "Battery out-gassing due to battery over charging and/or secondary failure mode of deep-discharge". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0718-03
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
KCI USA, Inc.
Product
TransportAir, Accessory to an Air Bed, Auxiliary Blower Unit, manufactured by Kinetic Concepts, Inc.
Product code
INX Bed, Air Fluidized
Reason for recall
Battery out-gassing due to battery over charging and/or secondary failure mode of deep-discharge.
FDA root cause
Other
Action
In approximately November 1997 the firm implemented the TransportAir retrograde for all active units.
Quantity in commerce
4264
Distribution
Nationwide
Product codes and lots
Serial numbers 0550037 through 0559399.