Z-0526-03
Philips BV300 Mobile X-Ray System
Philips North America LLC / initiated 2002-09-27 / Terminated
Philips Medical Systems began a recall of Philips BV300 Mobile X-Ray System on . The reason FDA records is "The C-bow support can bend and break allowing C-arm to fall from support". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0526-03
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- Philips BV300 Mobile X-Ray System
- Product code
- IZL System, X-Ray, Mobile
- Reason for recall
- The C-bow support can bend and break allowing C-arm to fall from support
- FDA root cause
- Other
- Action
- The firm issued a letter dated 9/27/02 to their customers. The letter advised of the problem and that a service representative will visit to inspect unit and correct.
- Quantity in commerce
- 229 devices
- Distribution
- The firm distributed to hospitals, x-ray centers and medical centers throughout the U.S.
- Product codes and lots
- Part numbers 9896 000 06801 and 9806 000 06802 with serial numbers: DBXXXX - all DBXXXX - all DDXXXX - all DEXXXX - all DG0001-DG0129 DH0001-DH0033 DJ0001 - DJ0152 DK0001 - DK 0076
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.
