Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0526-03

Philips BV300 Mobile X-Ray System

Philips North America LLC / initiated 2002-09-27 / Terminated

Philips Medical Systems began a recall of Philips BV300 Mobile X-Ray System on . The reason FDA records is "The C-bow support can bend and break allowing C-arm to fall from support". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0526-03
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
Philips BV300 Mobile X-Ray System
Product code
IZL System, X-Ray, Mobile
Reason for recall
The C-bow support can bend and break allowing C-arm to fall from support
FDA root cause
Other
Action
The firm issued a letter dated 9/27/02 to their customers. The letter advised of the problem and that a service representative will visit to inspect unit and correct.
Quantity in commerce
229 devices
Distribution
The firm distributed to hospitals, x-ray centers and medical centers throughout the U.S.
Product codes and lots
Part numbers 9896 000 06801 and 9806 000 06802 with serial numbers: DBXXXX - all DBXXXX - all DDXXXX - all DEXXXX - all DG0001-DG0129 DH0001-DH0033 DJ0001 - DJ0152 DK0001 - DK 0076