Z-0469-03
Megadyne MegaTip, Modified E-Z Clean Non Stick Cautery Tip, CAT/REF 0600M, Single Use Only, STERILE
Megadyne Medical Products, Inc / initiated 2002-12-20 / Terminated
Megadyne Medical Products, Inc began a recall of Megadyne MegaTip, Modified E-Z Clean Non Stick Cautery Tip, CAT/REF 0600M, Single Use Only, STERILE on . The reason FDA records is "The metal substrate cautery tip may break during use". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0469-03
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Megadyne Medical Products, Inc
- Product
- Megadyne MegaTip, Modified E-Z Clean Non Stick Cautery Tip, CAT/REF 0600M, Single Use Only, STERILE
- Product code
- GEI Electrosurgical, Cutting & Coagulation & Accessories
- Reason for recall
- The metal substrate cautery tip may break during use.
- FDA root cause
- Other
- Action
- On 12/20/2002, all consignees were notified by letter to return all affected product.
- Quantity in commerce
- 4855
- Distribution
- Nationwide. Foreign distribution to Australia, Belgium, Canada, Italy, Korea, New Zealand, Sweden.
- Product codes and lots
- Catalog Number 0600M. Lots 00994, 01880, 01181, 01112, 10156, 10670, 11008, 11388, 20260, 20814.
FIRST ARTICLE / WEEKLY