Z-0465-03
Premier Platinum brand HpSA, diagnostic test kits, one kit per pouch, one pouch per shipping box
Meridian Bioscience Inc / initiated 2002-12-26 / Terminated / root cause: labeling and UDI
Meridian Bioscience Inc began a recall of Premier Platinum brand HpSA, diagnostic test kits, one kit per pouch, one pouch per shipping box on . The reason FDA records is "Due to defective packaging, the firm cannot assure product stability through the labeled expiration date". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0465-03
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Meridian Bioscience Inc
- Product
- Premier Platinum brand HpSA, diagnostic test kits, one kit per pouch, one pouch per shipping box.
- Product code
- LYR Helicobacter Pylori
- Reason for recall
- Due to defective packaging, the firm cannot assure product stability through the labeled expiration date.
- Root cause tag
- labeling and UDI
- FDA root cause
- Other
- Action
- The firm mailed notices to their affected customers on 12/26/2002. The consignees were instructed to visually and physically inspect the seals on each microwell pouch for gaps. If properly sealed, the use of the microwells may be continued. Test wells in pouches with defective seals are to be quarantined and destroyed. The customers are asked to report back to the firm. All test microwell plates in defective pouches will be replaced by the firm.
- Quantity in commerce
- 1,020 kits
- Distribution
- The kits were distributed to end-users and distributors located nationwide; and to foreign distributors located in China, Guatemala, New Zealand, Hong Kong, Mexico, Great Britain, Belgium, Italy, Brazil, Australia, Canada, Thailand, Honduras, and Japan.
- Product codes and lots
- Lot #601348.069, Exp Date 22 SEP 03; Lot #601348.070, Exp Date 22 SEP 03; Kit Catalog #601348