Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0628-03

BIOMERIEUX, Thromboplastin Reagent, Simplastin Excel S, 6 ml Diluent, bioMerieux, Inc., Durham, NC 27704

bioMerieux, Inc. / initiated 2002-12-23 / Terminated

Biomerieux, Inc. began a recall of BIOMERIEUX, Thromboplastin Reagent, Simplastin Excel S, 6 ml Diluent, bioMerieux, Inc., Durham, NC 27704 on . The reason FDA records is "Vial to vial variability which can result in prolonged clot times". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0628-03
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
bioMerieux, Inc.
Product
BIOMERIEUX, Thromboplastin Reagent, Simplastin Excel S, 6 ml Diluent, bioMerieux, Inc., Durham, NC 27704
Product code
GJS Test, Time, Prothrombin
Reason for recall
Vial to vial variability which can result in prolonged clot times.
FDA root cause
Other
Action
Consignees were notified by letter via regular mail on December 23, 2002. They were advised to cease using the product, pull affected lot numbers, destroy the material and contact bioMerieux for replacement. Expanded recall 3/18/2003, consignees were notified by letter on March 19th , and March 31, 2003. Consignees were advised to stop using the affected lot numbers, destroy the remaining material in inventory and to contact their local bioMerieux Customer Service representative for replacement.
Quantity in commerce
4276 kits of lot #161553, 4275 kits of lot #161580, 4279 kits of lot #161581 and 4300 kits of lot 161661. Additional 2,280 kits of lot 161662.
Distribution
CA, CO, FL, GA, IA, IN, MN, NC, NJ, TN, WI. Australia, Brazil, Buenos Aires, Canada, Colombia, Hong Kong, Japan, Korea, Mexico, Netherlands, Thailand and the United Kingdom. There were no government accounts involved.
Product codes and lots
Product# 252182- Lot #161553, 161580, 161581 and 161661. Additional Catalog number: 252183, Lot number: 161662. Additional lot numbers recalled 3/19 & 31/2003: Lot #161639, 161664, 161716 and 161717. Recall expanded on 5/30/2003 to include the balance of the lot number produced under previous manufacturing practice. Product Number 252182 (6 ml), Lot number 161506; Prduct Number 252183 (20 ml), Lot Number 161503, 161504, 161554, 161582 and 161640.