Z-0556-03

ThermaCool TC 1.0 cm2 Treatment Tip, Single Patient Use Only

Solta Medical, Inc. / initiated 2002-12-09 / Terminated

Thermage began a recall of ThermaCool TC 1.0 cm2 Treatment Tip, Single Patient Use Only on . The reason FDA records is "Product tip has potential to liquid leakage into the internal circuitry causing malfunction and may result in epidermal injury". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0556-03
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Solta Medical, Inc.
Product
ThermaCool TC 1.0 cm2 Treatment Tip, Single Patient Use Only
Product code
GEI Electrosurgical, Cutting & Coagulation & Accessories
Reason for recall
Product tip has potential to liquid leakage into the internal circuitry causing malfunction and may result in epidermal injury.
FDA root cause
Other
Action
The recall notification was faxed to all 54 consignees on 12/8/02. A follow-up hard copy of the recall notification was sent via FedEx on 12/9/02.
Quantity in commerce
1,326 units
Distribution
The product was distributed nationwide. A total of 54 consignees received the product. The recall was appropriately extended to the user level; i.e., the wholesalers/distributors, hospitals and medical supply distributors, and physicians who received the recalled product. There is no known U. S. Government or Canadian distribution.
Product codes and lots
M00023,  M00024,  M00029, M00031, M00033R, M00034, M00035,