Z-0597-03
Intellivue Information Center Model: M3150BU#CO1
Philips North America LLC / initiated 2003-01-02 / Terminated
Philips Medical Systems, Inc. Cardiac & Monitoring Systems began a recall of Intellivue Information Center Model: M3150BU#CO1 on . The reason FDA records is "Central Station Monitoring Information Centers may experience sreeen freeze, i.e. loss of display, keyboard or mouse lock up". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0597-03
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- Intellivue Information Center Model: M3150BU#CO1
- Reason for recall
- Central Station Monitoring Information Centers may experience sreeen freeze, i.e. loss of display, keyboard or mouse lock up
- FDA root cause
- Other
- Action
- Philips Medical Systems notified users by Certified Mail on/about 2/19/03 that the firm was initiating a device field correction (recall) to conduct an upgrade to units witha HP VL400 or VL420 computer which is cconfigured for dual display by replacing the video and network communication card. Previous to this notification, Philips sent letters on 1/2/03 addressed to Biomedical Engineering Manager to advise users to reboot the system in the event of a system freeze.
- Quantity in commerce
- 7 units
- Distribution
- Nationwide Canada, Europe, Asia, Latin America
- Product codes and lots
- 4225A00169 4225A00171 4225A00181 4225A00182 4213A13456 4131A13457