Z-0613-03
REMEL CXM 30 Cefuroxime 30 mcg, Antimicrobial Susceptibility Test Disk, 50 disks to a cartridge, 5 cartridges...
Remel Inc / initiated 2002-12-05 / Terminated / root cause: packaging
Remel, Inc. began a recall of REMEL CXM 30 Cefuroxime 30 mcg, Antimicrobial Susceptibility Test Disk, 50 disks to a cartridge, 5 cartridges to a kit, catalog Number 33999. The label shows... on . The reason FDA records is "Packages of Cefuroxime susceptibility disks erroneously contained Cefaclor susceptibility disks". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-0613-03
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Remel Inc
- Product
- REMEL CXM 30 Cefuroxime 30 mcg, Antimicrobial Susceptibility Test Disk, 50 disks to a cartridge, 5 cartridges to a kit, catalog Number 33999. The label shows the product is ''Mfg for'' REMEL, Lenexa,KS.
- Product code
- JTN Susceptibility Test Discs, Antimicrobial
- Reason for recall
- Packages of Cefuroxime susceptibility disks erroneously contained Cefaclor susceptibility disks.
- Root cause tag
- packaging
- FDA root cause
- Other
- Action
- Recall letters dated 12/5/02 were issued requesting destruction of the product by the customer.
- Quantity in commerce
- 104 Pkg
- Distribution
- The product was distributed nationwide to 62 medical facilities including 3 military and 4 government accounts. No foreign accounts were reported.
- Product codes and lots
- Lot Number: 254243 - Exp. 09/04