Z-0546-03

Palacos R radiopaque bone cement, 40 gram. Product number 424800

Zimmer GmbH / initiated 2003-01-17 / Terminated / root cause: packaging

Biomet Inc began a recall of Palacos R radiopaque bone cement, 40 gram. Product number 424800 on . The reason FDA records is "There may be cracks in the blister pack overwrap for the glass ampoule containing the liquid monomer". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-0546-03
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
Palacos R radiopaque bone cement, 40 gram. Product number 424800.
Product code
LOD Bone Cement
Reason for recall
There may be cracks in the blister pack overwrap for the glass ampoule containing the liquid monomer.
Root cause tag
packaging
FDA root cause
Other
Action
The firm issued recall letter for one lot dated January 21, 2003, but a subsequent letter dated January 21, 2003 expanded the recall to all lots. Customers are instructed to not use the product. They are to either return it to their local distributor, or to mix the cement and properly dispose of it themselves. [This is a hazardous material requiring special shipping methods.]
Quantity in commerce
301,226 units
Distribution
United States
Product codes and lots
All lots.