Z-0590-03
Scimed ChoICE PT Extra Support PTCA Guide Wire with ICE Hydrophilic Coating, 300 cm length
Boston Scientific Corporation / initiated 2002-08-22 / Terminated / root cause: labeling and UDI
Symbiosis Corp. began a recall of Scimed ChoICE PT Extra Support PTCA Guide Wire with ICE Hydrophilic Coating, 300 cm length on . The reason FDA records is "Product labeled as Scimed ChoICE PT Extra Suport PTCA Guide Wire 300 cm were actually Scimed PT Graphix Intermediate PTCA Guide Wire 300 cm". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0590-03
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Boston Scientific Corporation
- Product
- Scimed ChoICE PT Extra Support PTCA Guide Wire with ICE Hydrophilic Coating, 300 cm length
- Product code
- DQX Wire, Guide, Catheter
- Reason for recall
- Product labeled as Scimed ChoICE PT Extra Suport PTCA Guide Wire 300 cm were actually Scimed PT Graphix Intermediate PTCA Guide Wire 300 cm
- Root cause tag
- labeling and UDI
- FDA root cause
- Other
- Action
- Recall letters with response forms were sent to hosptials via FED EX on 8/22/02. Product was to be returned to Boston Scientific in Quincy, MA.
- Quantity in commerce
- 27/5 pack boxes
- Distribution
- Recalled product was shipped to 107 domestic hosptials nationwide and to 2 international acccounts.
- Product codes and lots
- Individual Devices in Pouches with Catalog No. 12155-01, UPN H74912155011, Lot Number 4773987, exp. date 2004-5. These pouches were packaged in two lots of five-pack boxes with the following codes: Catalog 12155-01, lot 4791304 Catalog 12155-01, lot 4798294