Z-0608-03

Compat 500 mL Baggle Top Fill Enteral Feeding Container with Preattached Enteral Delivery Pump Set, Item...

Novartis Nutrition Corp. / initiated 2003-01-27 / Terminated

Novartis Nutrition Corp began a recall of Compat 500 mL Baggle Top Fill Enteral Feeding Container with Preattached Enteral Delivery Pump Set, Item number 19931200 on . The reason FDA records is "Some of the sets may leak or disconnect near the pump adapter during use. The patient could ingest air and vomit causing a safety concern". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0608-03
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Novartis Nutrition Corp.
Product
Compat 500 mL Baggle Top Fill Enteral Feeding Container with Preattached Enteral Delivery Pump Set, Item number 19931200
Product code
LZH Pump, Infusion, Enteral
Reason for recall
Some of the sets may leak or disconnect near the pump adapter during use. The patient could ingest air and vomit causing a safety concern.
FDA root cause
Other
Action
The recall letter, dated January 31, 2003, requested that the consignees return the recalled products and notify 'end users' to stop using the products.
Quantity in commerce
803 cases (30 tube sets per case)
Distribution
Nationwide in the United States and to Canada
Product codes and lots
Item number 19931200, lots 3152A and 3182A