Z-0727-03
ICON 9.5 Software used on the ICON workstations designed for use with the E.CAM Emission Imaging Computed...
Siemens Medical Solutions USA, Inc. / initiated 2003-02-10 / Terminated / root cause: software
Siemens Medical Solutions USA, Inc. began a recall of ICON 9.5 Software used on the ICON workstations designed for use with the E.CAM Emission Imaging Computed Tomography Systems; Siemens Medical Systems, Inc... on . The reason FDA records is "ICON 9.5 software does not apply flood corrections during whole body SPECT studies with multiple bed positions". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0727-03
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- ICON 9.5 Software used on the ICON workstations designed for use with the E.CAM Emission Imaging Computed Tomography Systems; Siemens Medical Systems, Inc., Nuclear Medicine Group, Hoffman Estates, IL 60195-5203
- Product code
- JWM System, Tomographic, Nuclear
- Reason for recall
- ICON 9.5 software does not apply flood corrections during whole body SPECT studies with multiple bed positions.
- Root cause tag
- software
- FDA root cause
- Other
- Action
- Siemens sent Customer Advisory letters dated 2/10/03 to all ICON 9.5 customers with Whole Body SPECT functionality on integrated camera systems including Diacam, Multispect 2, Multispect 3 and e.cam camera systems aquired on an ICON workstation. The accounts were informed of the problem related to applying the flood correction matrix to Whole Body SPECT studies in ICON version 9.5 software, and were provided the follwoing workaround: To ensure that the flood correction is applied, the user should acquire multiple single bed positions with the General SPECT predefine, indexing the bed manually between each SPECT acquisition. The data can then be combined into a Whole Body SPECT in the Whole Body SPECT Display program for review. In addition, users were requested to routinely perform standard quality control procedures at the facility to ensure proper image quality. No software upgrade is planned.
- Quantity in commerce
- 600 units
- Distribution
- Nationwide, United Arab Emirates, Argentina, Austria, Australia, Barbados, Belgium, Brazil, Canada, Chile, China, Colombia, Czech Republic, Equador, Egypt, France, Germany, Great Britain, Georgia, Greece, Hong Kong, Israel, India, Italy, Jordan, Japan, Korea, Latvia, Mexico, Niger, Norway, New Zealand, Philippines, Pakistan, Poland, Russia, Saudi Arabia, SIngapore, Spain, Sweden, Switzerland, Taiwan, Turkey, Ukraine and the Netherlands.
- Product codes and lots
- All Diacam, Multispect 2, Multispect 3 and e.cam camera systems with ICON workstations with ICON software version 9.5 and the optional Whole Body SPECT software.