Z-0715-03
Accu-Chek Comfort Curve glucose test strips; part numbers 2030420, 2030365, 2030373, 2030381, 3000133 and...
Roche Diagnostics GmbH / initiated 2003-02-12 / Terminated
Roche Diagnostics Corp. began a recall of Accu-Chek Comfort Curve glucose test strips; part numbers 2030420, 2030365, 2030373, 2030381, 3000133 and 3000141 on . The reason FDA records is "Crack in the bottom of the test strip vial will cause erroneously low or high blood glucose readings". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0715-03
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Roche Diagnostics GmbH
- Product
- Accu-Chek Comfort Curve glucose test strips; part numbers 2030420, 2030365, 2030373, 2030381, 3000133 and 3000141.
- Product code
- LFR Glucose Dehydrogenase, Glucose
- Reason for recall
- Crack in the bottom of the test strip vial will cause erroneously low or high blood glucose readings.
- FDA root cause
- Other
- Action
- Urgent Product Correction letters dated 2/12/03 were issued to pharmacists, specific health care providers, and the laboratory manager at customer accounts. Letter recipients were asked to advise their patients to inspect their vial for a crack prior to each use.
- Distribution
- United States, Argentina, Australia, Brazil, Canada, Chile, China, Colombia, Denmark, Finland, France, Germany, Guatemala, Honduras, Hong Kong, Indonesia, Japan, Malaysia, Mexico, New Zealand, Norway, Pakistan, Peru, Philippines, Portugal, Singapore, South Africa, South Korea, Spain, Taiwan, Thailand, United Kingdom and Vietnam.
- Product codes and lots
- All lots.