Z-0716-03
Accu-Chek Advantage glucose test strips; part numbers 336, 556, 787 and 966
Roche Diagnostics GmbH / initiated 2003-02-12 / Terminated
Roche Diagnostics Corp. began a recall of Accu-Chek Advantage glucose test strips; part numbers 336, 556, 787 and 966 on . The reason FDA records is "Crack in the bottom of the test strip vial will cause erroneously low or high blood glucose readings". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0716-03
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Roche Diagnostics GmbH
- Product
- Accu-Chek Advantage glucose test strips; part numbers 336, 556, 787 and 966.
- Product code
- LFR Glucose Dehydrogenase, Glucose
- Reason for recall
- Crack in the bottom of the test strip vial will cause erroneously low or high blood glucose readings.
- FDA root cause
- Other
- Action
- Urgent Product Correction letters dated 2/12/03 were issued to pharmacists, specific health care providers, and the laboratory manager at customer accounts. Letter recipients were asked to advise their patients to inspect their vial for a crack prior to each use.
- Distribution
- United States, Argentina, Australia, Brazil, Canada, Chile, China, Colombia, Denmark, Finland, France, Germany, Guatemala, Honduras, Hong Kong, Indonesia, Japan, Malaysia, Mexico, New Zealand, Norway, Pakistan, Peru, Philippines, Portugal, Singapore, South Africa, South Korea, Spain, Taiwan, Thailand, United Kingdom and Vietnam.
- Product codes and lots
- All lots.
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.
