Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0617-03

Integrated Orbiter Single Detector General Purpose Gamma Camera System; Siemens Medical Systems, Inc...

Siemens Medical Solutions USA, Inc. / initiated 2003-02-10 / Terminated

Siemens Medical Solutions USA, Inc. began a recall of Integrated Orbiter Single Detector General Purpose Gamma Camera System; Siemens Medical Systems, Inc., Nuclear Medicine Group, 2501 North Barrington Road... on . The reason FDA records is "Unintended yoke rotational motion". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0617-03
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Integrated Orbiter Single Detector General Purpose Gamma Camera System; Siemens Medical Systems, Inc., Nuclear Medicine Group, 2501 North Barrington Road, Hoffman Estates, IL 60195-5203
Product code
IYX Camera, Scintillation (Gamma)
Reason for recall
Unintended yoke rotational motion.
FDA root cause
Other
Action
Siemens sent Customer Advisory letters dated 2/10/03 to all Integrated Orbiter customers. The accounts were informed of the problem with unexpected camera movement in the opposite direction at the beginning of a SPECT study, and were advised to use the emergency stop at any indication of improper camera movement. The customers were advised that they would be contacted by their Siemens Customer Service Representative to schedule a firmware upgrade that will prevent the occurrence of unintended camera rotation. The upgrade will be done free of charge.
Quantity in commerce
256 units
Distribution
Nationwide, United Arab Emirates, Argentina, Austria, Azerbaizan,Bangladesh, Belgium, Brazil, Canada, China, Colombia, Cyprus, France, Germany, Great Britain, Greece, Hong Kong, India, Ireland, Italy, Jordan, Japan, Korea, Norway, Pakistan, Romania, Switzerland, Tunisia, Venezuela and the Netherlands.
Product codes and lots
model 35-26-932, serial numbers 102, 1001/1217, 1169A; model 35-44-637, serial numbers 0101/0142, 120A