Z-0617-03
Integrated Orbiter Single Detector General Purpose Gamma Camera System; Siemens Medical Systems, Inc...
Siemens Medical Solutions USA, Inc. / initiated 2003-02-10 / Terminated
Siemens Medical Solutions USA, Inc. began a recall of Integrated Orbiter Single Detector General Purpose Gamma Camera System; Siemens Medical Systems, Inc., Nuclear Medicine Group, 2501 North Barrington Road... on . The reason FDA records is "Unintended yoke rotational motion". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0617-03
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- Integrated Orbiter Single Detector General Purpose Gamma Camera System; Siemens Medical Systems, Inc., Nuclear Medicine Group, 2501 North Barrington Road, Hoffman Estates, IL 60195-5203
- Product code
- IYX Camera, Scintillation (Gamma)
- Reason for recall
- Unintended yoke rotational motion.
- FDA root cause
- Other
- Action
- Siemens sent Customer Advisory letters dated 2/10/03 to all Integrated Orbiter customers. The accounts were informed of the problem with unexpected camera movement in the opposite direction at the beginning of a SPECT study, and were advised to use the emergency stop at any indication of improper camera movement. The customers were advised that they would be contacted by their Siemens Customer Service Representative to schedule a firmware upgrade that will prevent the occurrence of unintended camera rotation. The upgrade will be done free of charge.
- Quantity in commerce
- 256 units
- Distribution
- Nationwide, United Arab Emirates, Argentina, Austria, Azerbaizan,Bangladesh, Belgium, Brazil, Canada, China, Colombia, Cyprus, France, Germany, Great Britain, Greece, Hong Kong, India, Ireland, Italy, Jordan, Japan, Korea, Norway, Pakistan, Romania, Switzerland, Tunisia, Venezuela and the Netherlands.
- Product codes and lots
- model 35-26-932, serial numbers 102, 1001/1217, 1169A; model 35-44-637, serial numbers 0101/0142, 120A