Z-0623-03
e.cam Emission Imaging Computed Tomography Systems used for all common nuclear medicine procedures; Siemens...
Siemens Medical Solutions USA, Inc. / initiated 2003-02-10 / Terminated
Siemens Medical Solutions USA, Inc. began a recall of e.cam Emission Imaging Computed Tomography Systems used for all common nuclear medicine procedures; Siemens Medical Systems, Inc., Nuclear Medicine Group... on . The reason FDA records is "Unexpected release of the radial and rotate brakes in the case of multiple simultaneous fault conditions". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0623-03
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- e.cam Emission Imaging Computed Tomography Systems used for all common nuclear medicine procedures; Siemens Medical Systems, Inc., Nuclear Medicine Group, Hoffman Estates, IL 60195-5203
- Product code
- KPS System, Tomography, Computed, Emission
- Reason for recall
- Unexpected release of the radial and rotate brakes in the case of multiple simultaneous fault conditions.
- FDA root cause
- Other
- Action
- Siemens sent Customer Advisory letters dated 2/10/03 to all e.cam customers. The accounts were informed of the problem with unexpected camera movement when multiple fault cocnditions are presnet, causing the brake system to release and the roation of the detector assemblies, and were advised to use the emergency stop at any indication of improper camera movement. The customers were advised that they would be contacted by their Siemens Customer Service Representative to schedule a firmware upgrade that will prevent the occurrence of unintended release of the radial and rotate brakes. The upgrade will be done free of charge.
- Quantity in commerce
- 1927 units
- Distribution
- Nationwide, United Arab Emirates, Argentina, Austria, Australia, Bosnia, Barbados, Bangladesh, Belgium, Bolivia, Brazil, Camaroon, Canada, Chile, China, Colombia, Republic of Congo, Czech Republic, Denmark, Dominican Republic, El Salvador, Equador, Egypt, France, Finland, Germany, Great Britain, Georgia, Greece, Hong Kong, Honduras, Ireland, Israel, India, Italy, Iraq, Jordan, Japan, Korea, Kuwait, Latvia, Lebanon, Lithuania, Luxembourg, Mexico, Morocco, Madagascar, Myanmar, Mauritius, Malaysia, Niger, Norway, New Zealand, Oman, Peru, Philippines, Pakistan, Poland, Potugal, Russia, Saudi Arabia, Singapore, Spain, Sweden, Switzerland, Taiwan, Turkey, Ukraine, Uzbekistan, South Africa and the Netherlands.
- Product codes and lots
- All e.cam towers, serial numbers 00013 through 08543, utilizing e.soft, ICON-S and duet acquisition systems, and e.soft and ICON-S COIN systems.