Z-0669-03
LH 700 Series Hematology Analyzers. Part numbers 66056032, 6605633, 6605645
Siemens Healthcare Diagnostics Inc. / initiated 2002-12-27 / Terminated / root cause: software
Beckman Coulter Inc began a recall of LH 700 Series Hematology Analyzers. Part numbers 66056032, 6605633, 6605645 on . The reason FDA records is "Incorrect Hemoglobin result can be reported with software version 2A and higher". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0669-03
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- LH 700 Series Hematology Analyzers. Part numbers 66056032, 6605633, 6605645
- Product code
- GKZ Counter, Differential Cell
- Reason for recall
- Incorrect Hemoglobin result can be reported with software version 2A and higher.
- Root cause tag
- software
- FDA root cause
- Other
- Action
- Customers were contacted by telephone and mail from 1/27/03 to 2/3/2003. Notice included modified operating instructions. A software fix is being developed and will be installed by Beckman personnel.
- Quantity in commerce
- 234
- Distribution
- Nationwide and Canada.
- Product codes and lots
- All instruments with version 2A or higher.