Z-0670-03

Misys Laboratory System Version 5.3

CliniSys, Inc / initiated 2003-02-19 / Terminated / root cause: software

Misys Healthcare Systems began a recall of Misys Laboratory System Version 5.3 on . The reason FDA records is "Software anomaly. The HIDE command used to suppress suspected individual microbiology observations does not work in GUI inquiry". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0670-03
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
CliniSys, Inc
Product
Misys Laboratory System Version 5.3
Product code
JQP Calculator/Data Processing Module, For Clinical Use
Reason for recall
Software anomaly. The HIDE command used to suppress suspected individual microbiology observations does not work in GUI inquiry
Root cause tag
software
FDA root cause
Other
Action
Notification was made by FAX to customers. A procedureal work around was provided also. The firm will announce the availability of a software fix in the future.
Quantity in commerce
464
Distribution
Nationwide and to United Kingdom, Ireland, Canada, Denmark, Bermuda, Saudi Arabia, United Arab Emirates.
Product codes and lots
Version 5.3