Abbott IMx AFP Calibrators, list 2271-01; a box of 6 bottles (4 mL each) with AFP (human), nonreactive for...
Abbott Laboratories HPD/ADD began a recall of Abbott IMx AFP Calibrators, list 2271-01; a box of 6 bottles (4 mL each) with AFP (human), nonreactive for anti-HIV-1/HIV-2 and HCV and nonreactive for HBsAg... on . The reason FDA records is "The IMx AFP Calibrators may cause depressed values for the IMx AFP Low Control, and cause patient results to be 20% lower than actual values". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0726-03
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Medical
- Product
- Abbott IMx AFP Calibrators, list 2271-01; a box of 6 bottles (4 mL each) with AFP (human), nonreactive for anti-HIV-1/HIV-2 and HCV and nonreactive for HBsAg, inbuffered calf serum at the following concentrations: A - 0, B - 15, C - 50, D - 100, E - 200 and F - 350; Abbott Diagnostics, a Division of Abbott Laboratories, Abbott park, IL 60064
- Product code
- LOJ Kit, Test,Alpha-Fetoprotein For Testicular Cancer
- Reason for recall
- The IMx AFP Calibrators may cause depressed values for the IMx AFP Low Control, and cause patient results to be 20% lower than actual values
- Root cause tag
- manufacturing process
- FDA root cause
- Other
- Action
- Recalled by letter dated 3/4/03. The laboratory accounts were informed of the lot of calibrators causing depressed values for the IMx AFP Low Control, requested to destroy any of the suspect lot of calibrators, and recalibrate the AFP assay using a different lot of calibrators.
- Quantity in commerce
- 107 kits
- Distribution
- Nationwide
- Product codes and lots
- lot number 89651M300, exp. 04/20/03