Z-0681-03

IMx CA 15-3 Reagent Pack, list 6A75-22; this reagent pack contains one 4-mL bottle of Mode 1 Calibrator along...

Abbott Medical / initiated 2003-03-04 / Terminated / root cause: manufacturing process

Abbott Laboratories HPD/ADD began a recall of IMx CA 15-3 Reagent Pack, list 6A75-22; this reagent pack contains one 4-mL bottle of Mode 1 Calibrator along with other reagents; Abbott Laboratories, Abbott... on . The reason FDA records is "The CA 15-3 Calibrators may cause depressed values for the CA 15-3 High Control and for patient specimens". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0681-03
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
IMx CA 15-3 Reagent Pack, list 6A75-22; this reagent pack contains one 4-mL bottle of Mode 1 Calibrator along with other reagents; Abbott Laboratories, Abbott Park, IL
Product code
MOI System, Test, Immunological, Antigen, Tumor
Reason for recall
The CA 15-3 Calibrators may cause depressed values for the CA 15-3 High Control and for patient specimens.
Root cause tag
manufacturing process
FDA root cause
Other
Action
Recalled by letter dated 3/4/03. The accounts were informed of the depressed values for the CA 15-3 High Control and for patient specimens, and were requested to discontinue use and destroy any of the three affected lots found in inventory.
Quantity in commerce
409 packs
Distribution
Nationwide and Ecuador, England, Hong Kong, Chile, Japan, Canada, Singapore, Australia, Pakistan, Argentina, Thailand, Aruba, Korea and Jamaica.
Product codes and lots
list 6A75-22, lot 90843M200