IMx CA 15-3 Reagent Pack, list 6A75-22; this reagent pack contains one 4-mL bottle of Mode 1 Calibrator along...
Abbott Laboratories HPD/ADD began a recall of IMx CA 15-3 Reagent Pack, list 6A75-22; this reagent pack contains one 4-mL bottle of Mode 1 Calibrator along with other reagents; Abbott Laboratories, Abbott... on . The reason FDA records is "The CA 15-3 Calibrators may cause depressed values for the CA 15-3 High Control and for patient specimens". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0681-03
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Medical
- Product
- IMx CA 15-3 Reagent Pack, list 6A75-22; this reagent pack contains one 4-mL bottle of Mode 1 Calibrator along with other reagents; Abbott Laboratories, Abbott Park, IL
- Product code
- MOI System, Test, Immunological, Antigen, Tumor
- Reason for recall
- The CA 15-3 Calibrators may cause depressed values for the CA 15-3 High Control and for patient specimens.
- Root cause tag
- manufacturing process
- FDA root cause
- Other
- Action
- Recalled by letter dated 3/4/03. The accounts were informed of the depressed values for the CA 15-3 High Control and for patient specimens, and were requested to discontinue use and destroy any of the three affected lots found in inventory.
- Quantity in commerce
- 409 packs
- Distribution
- Nationwide and Ecuador, England, Hong Kong, Chile, Japan, Canada, Singapore, Australia, Pakistan, Argentina, Thailand, Aruba, Korea and Jamaica.
- Product codes and lots
- list 6A75-22, lot 90843M200