Z-0684-03
Ethanol Standard, 0.5mg Ethanol in 1 mL. Product size configuration: 1 mL glass ampule shipped 10 ampules per...
EMD Millipore Corporation / initiated 2002-12-13 / Terminated
EMD Chemicals Inc began a recall of Ethanol Standard, 0.5mg Ethanol in 1 mL. Product size configuration: 1 mL glass ampule shipped 10 ampules per case. Item Number 68988/95 on . The reason FDA records is "The Ethanol Standard is OOS on the low end. The results were 0.44 mg/mL while the specification is 0.48-0.52 mg/mL". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0684-03
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- EMD Millipore Corporation
- Product
- Ethanol Standard, 0.5mg Ethanol in 1 mL. Product size configuration: 1 mL glass ampule shipped 10 ampules per case. Item Number 68988/95
- Product code
- LCH Tonometer (Calibration And Q.C. Of Blood-Gas Instruments), Clinical
- Reason for recall
- The Ethanol Standard is OOS on the low end. The results were 0.44 mg/mL while the specification is 0.48-0.52 mg/mL.
- FDA root cause
- Other
- Action
- All customers were nofitied by phone and letter. The first notification occurred on 12/17/2002 followed with additional notifications on 1/24/2003, 1/28/2003, and 2/4/2003.
- Quantity in commerce
- 139 cases of 10 ampules
- Distribution
- The Ethanol Products were distributed nationwide by the following distributors: VWR International, 1310 Goshen Pkwy, West Chester, PA 19380; Government Scientific, 8460 K-Tyco Road, Vienna, VA 22182; Allegiance Healthcare, 1430 Waukegan Rd., McGaw Park, IL 60085; Fisher Scientific, 2000 Park Lane, Pittsburgh, PA 15275; Hawaii Chemical Co., 2363 North King Street, Honolulu, HI 96819; Krackeler Scientific Inc., 57 Broadway, Albany, NY 12202; Midland Scientific Inc., 1202 South 11th St., Omaha, NE 68108; Reagents, Inc. 4746 Sweden Rd., Charlotte, NC 28224. The Papanicolaou Stain was distributed nationwide by the following distributors: VWR International, 1310 Goshen Pkwy, West Chester, PA 19380; Fisher Scientific, 2000 Park Lane, Pittsburgh, PA 15275; Kramer Scientific, 711 Executive Blvd, Valley Cottage, NY 10989; Laboratory Supply Company, 3069 Mercury Rd., Jacksonville, FL 32207; Allegiance Healthcare, 1430 Waukegan Rd., McGaw Park, IL 60085; Infolab, 3000 Greenhill Dr., Round Rock, TX 78664; Government Scientific, 8460 K-Tyco Road, Vienna, VA 22182.
- Product codes and lots
- Item #68988/95; Lot #2330; Exp. Date May 31, 2005
FIRST ARTICLE / WEEKLY