Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0720-03

CoaguChek PT test strips, sold in 12 pack

Roche Diagnostics GmbH / initiated 2003-02-20 / Terminated

Roche Diagnostics Corp. began a recall of CoaguChek PT test strips, sold in 12 pack on . The reason FDA records is "Discrepancy between INR and the % Quick displayed on the meter". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0720-03
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Roche Diagnostics GmbH
Product
CoaguChek PT test strips, sold in 12 pack.
Product code
JPA System, Multipurpose For In Vitro Coagulation Studies
Reason for recall
Discrepancy between INR and the % Quick displayed on the meter.
FDA root cause
Other
Action
Customers in Germany were notified of the recall, and asked to return this lot of product. While the product is sold in many countries, Germany is the only country affected, since their results are the only ones reported as % Quick.
Quantity in commerce
11412 containers
Distribution
Germany
Product codes and lots
lot 533