Z-0712-03
Gluteraldehyde Concentration Indicators. MetriTest 1.5% and MetriTest 1.8%, also OmniChek 1.5% and OmniCheck...
Sybron Dental Specialties, Inc. / initiated 2003-02-25 / Terminated / root cause: packaging
Sybron Dental Specialties began a recall of Gluteraldehyde Concentration Indicators. MetriTest 1.5% and MetriTest 1.8%, also OmniChek 1.5% and OmniCheck 1.8% on . The reason FDA records is "Moisture ingress into packaging of test strips may cause premature failure". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-0712-03
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Sybron Dental Specialties, Inc.
- Product
- Gluteraldehyde Concentration Indicators. MetriTest 1.5% and MetriTest 1.8%, also OmniChek 1.5% and OmniCheck 1.8%
- Product code
- JOJ Indicator, Physical/Chemical Sterilization Process
- Reason for recall
- Moisture ingress into packaging of test strips may cause premature failure.
- Root cause tag
- packaging
- FDA root cause
- Other
- Action
- Notice was sent to customers of the potential problem on 3/7/2003. The notice included directions to perform a Modified Quality Control Procedure. Instructions are to test strips function each day.
- Quantity in commerce
- 4,112 cases of two bottles each. 60 strips per bottle.
- Distribution
- Nationwide and Canada.
- Product codes and lots
- MetriTest 1.5%, part number 10-303, lot numbers: 007917, 007919, 008184, 008325, 008665, 009100. MetriTest 1.8%, part number 10-304, lot numbers: 009101, 009121, 009144, 009249. OmniChek G 1.5%, part number 10-313, lot 009091. OmniChek G 1.8%, part number 10-313, lot 009102.