Z-0712-03

Gluteraldehyde Concentration Indicators. MetriTest 1.5% and MetriTest 1.8%, also OmniChek 1.5% and OmniCheck...

Sybron Dental Specialties, Inc. / initiated 2003-02-25 / Terminated / root cause: packaging

Sybron Dental Specialties began a recall of Gluteraldehyde Concentration Indicators. MetriTest 1.5% and MetriTest 1.8%, also OmniChek 1.5% and OmniCheck 1.8% on . The reason FDA records is "Moisture ingress into packaging of test strips may cause premature failure". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-0712-03
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Sybron Dental Specialties, Inc.
Product
Gluteraldehyde Concentration Indicators. MetriTest 1.5% and MetriTest 1.8%, also OmniChek 1.5% and OmniCheck 1.8%
Product code
JOJ Indicator, Physical/Chemical Sterilization Process
Reason for recall
Moisture ingress into packaging of test strips may cause premature failure.
Root cause tag
packaging
FDA root cause
Other
Action
Notice was sent to customers of the potential problem on 3/7/2003. The notice included directions to perform a Modified Quality Control Procedure. Instructions are to test strips function each day.
Quantity in commerce
4,112 cases of two bottles each. 60 strips per bottle.
Distribution
Nationwide and Canada.
Product codes and lots
MetriTest 1.5%, part number 10-303, lot numbers: 007917, 007919, 008184, 008325, 008665, 009100. MetriTest 1.8%, part number 10-304, lot numbers: 009101, 009121, 009144, 009249.  OmniChek G 1.5%, part number 10-313, lot 009091. OmniChek G 1.8%, part number 10-313, lot 009102.