Z-0734-03

Boston Scientific/Microvasive Open End Ureteral Catheter (4 Fr x 70 cm) Product Number: 400-200

Boston Scientific Corporation / initiated 2003-03-17 / Terminated

Boston Scientific Corporation began a recall of Boston Scientific/Microvasive Open End Ureteral Catheter (4 Fr x 70 cm) Product Number: 400-200 on . The reason FDA records is "Warning flag may be detached from the tip protector on the catheter and result in the tip protector advancing through the catheter during use". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0734-03
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Boston Scientific Corporation
Product
Boston Scientific/Microvasive Open End Ureteral Catheter (4 Fr x 70 cm) Product Number: 400-200
Product code
GBM Catheter, Urethral
Reason for recall
Warning flag may be detached from the tip protector on the catheter and result in the tip protector advancing through the catheter during use
FDA root cause
Other
Action
Boston Scientific/Microvasive, Natick, MA notified domestic accounts by letter on 3/17/03. Users are requesed to check inventory and return product to Boston Scientific. A responsed form was requested to be completed. and returned by facsmile.
Quantity in commerce
12 boxes (20 units/box)
Distribution
Nationwide Foreign:Singapore, Japan, Korea, Canada, Africa
Product codes and lots
Lot Numbers: 5178407 5202076 5275909 5309468 5337024