Z-0833-03

Baxter System 1000 Single Patient Hemodialysis Delivery System; Model SYS1000; all series 1000 instruments...

Baxter Healthcare Corporation / initiated 2003-03-20 / Terminated

Baxter Healthcare Corp. began a recall of Baxter System 1000 Single Patient Hemodialysis Delivery System; Model SYS1000; all series 1000 instruments labeled as System 1000, AltraTouch 1000, Baxter Tina... on . The reason FDA records is "The dialysis treatment may be run without the bicarbonate due to an undetected bicarbonate pump stall following a sodium change". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0833-03
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Baxter Healthcare Corporation
Product
Baxter System 1000 Single Patient Hemodialysis Delivery System; Model SYS1000; all series 1000 instruments labeled as System 1000, AltraTouch 1000, Baxter Tina and Baxter Aurora; Baxter Healthcare Corporation, Deerfield, IL 60015 USA or Baxter Healthcare Corporation, McGaw Park, IL 60085 USA
Product code
FKP System, Dialysate Delivery, Single Patient
Reason for recall
The dialysis treatment may be run without the bicarbonate due to an undetected bicarbonate pump stall following a sodium change.
FDA root cause
Other
Action
Baxter sent recall letters dated 3/20/03 to the Medical Director and Hemodialysis Chief Technicians via first class mail on 3/20/03. The letters informed the accounts of the potential of running a treatment without bicarbonate due to an undetected bicarbonate pump stall following a sodium change, and provided instructions to change the B-A-alarm limits on the System 1000 Delivery Systems and confirm proper functionality after completing these changes. These changes will prevent undetected acid-only delivery in the event of a sodium level change
Quantity in commerce
25,014 units
Distribution
The devices were distributed nationwide and internationally to Baxter subsidiaries in Argentina, Belgium, Bolivia, Brazil, Canada, Chile, China, Ecuador, Egypt, El Salvador, Guatemala, Honduras, Hong Kong, India, Mexico, Philippines, Singapore, South Korea, Thailand, Turkey, United Kingdom, Venezuela and Vietnam.
Product codes and lots
Serial numbers 1001S to 1142S; 50001 to 52109; 01001 to 23764