Z-0903-03

Hill-Rom brand Century+ series bed

Hill-Rom Inc. / initiated 2003-04-24 / Terminated

Hill-Rom, Inc. began a recall of Hill-Rom brand Century+ series bed on . The reason FDA records is "Possible electrical shock hazard, as the power cord grounding pin may break off or become detached". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0903-03
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Hill-Rom Inc.
Product
Hill-Rom brand Century+ series bed
Product code
KNC Table, Obstetric (And Accessories)
Reason for recall
Possible electrical shock hazard, as the power cord grounding pin may break off or become detached.
FDA root cause
Other
Action
Letters dated April 24, 2003 were sent to each consignee instructing them to inspect the beds for power cords with non-flared ground pins, to cut plug off any such cords when found, and replace these plugs with new plugs with the flared ground pin. [The type of plug on the unit can be seen by the naked eye, as the plug is transparent plastic molding.]
Distribution
United States, Bahamas, Bermuda, Canada, Colombia, Costa Rica, Dominican Republic, Guam, Japan, Mexico, Panama, Saudi Arabia, South Korea, Taiwan and Venezuela
Product codes and lots
All units distributed between January 1, 1999 and July 1, 2002.