Hill-Rom, Inc. began a recall of Hill-Rom brand Century+ series bed on . The reason FDA records is "Possible electrical shock hazard, as the power cord grounding pin may break off or become detached". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0903-03
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Hill-Rom Inc.
- Product
- Hill-Rom brand Century+ series bed
- Product code
- KNC Table, Obstetric (And Accessories)
- Reason for recall
- Possible electrical shock hazard, as the power cord grounding pin may break off or become detached.
- FDA root cause
- Other
- Action
- Letters dated April 24, 2003 were sent to each consignee instructing them to inspect the beds for power cords with non-flared ground pins, to cut plug off any such cords when found, and replace these plugs with new plugs with the flared ground pin. [The type of plug on the unit can be seen by the naked eye, as the plug is transparent plastic molding.]
- Distribution
- United States, Bahamas, Bermuda, Canada, Colombia, Costa Rica, Dominican Republic, Guam, Japan, Mexico, Panama, Saudi Arabia, South Korea, Taiwan and Venezuela
- Product codes and lots
- All units distributed between January 1, 1999 and July 1, 2002.