Z-0782-03
Architect Progesterone Reagent Kit; list 6C26-27 (100 tests) and 6C26-22 (400 tests); Abbott Diagnostics, a...
Abbott Medical / initiated 2003-03-31 / Terminated
Abbott Laboratories HPD/ADD began a recall of Architect Progesterone Reagent Kit; list 6C26-27 (100 tests) and 6C26-22 (400 tests); Abbott Diagnostics, a Division of Abbott Laboratories, Abbott Park, IL... on . The reason FDA records is "Lot gives lower specimen results compared to results generated with other lots". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0782-03
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Medical
- Product
- Architect Progesterone Reagent Kit; list 6C26-27 (100 tests) and 6C26-22 (400 tests); Abbott Diagnostics, a Division of Abbott Laboratories, Abbott Park, IL 60064
- Product code
- JLS Radioimmunoassay, Progesterone
- Reason for recall
- Lot gives lower specimen results compared to results generated with other lots.
- FDA root cause
- Other
- Action
- Recalled by telephone calls and follow-up letters dated 3/31/03. The accounts were informed of the depressed patient specimen results, and were requested to discontinue use and destroy any of the three affected lots found in inventory. The accounts were advised to notify physicians of the lower patient results observed with these lots and suggest review of the patient results to determine any impact on patient management.
- Quantity in commerce
- 368 kits
- Distribution
- Nationwide, Canada, Columbia, England, Japan, Australia, New Zealand and Singapore
- Product codes and lots
- list 6C26-27, lot 95651M201 list 6C26-22, lots 95651M200, 95651M202