Z-1029-03
Magic View 1000U, version VE40A. Digital Image Communication System
Siemens Medical Solutions USA, Inc. / initiated 2003-02-28 / Terminated / root cause: software
Siemens Medical Solutions USA, Inc began a recall of Magic View 1000U, version VE40A. Digital Image Communication System on . The reason FDA records is "software issue. New examinations may not be saved properly and the examination may be lost". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1029-03
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- Magic View 1000U, version VE40A. Digital Image Communication System
- Reason for recall
- software issue. New examinations may not be saved properly and the examination may be lost.
- Root cause tag
- software
- FDA root cause
- Other
- Action
- The recalling firm has issued a letter to their field service representative dated 2/28/02. The letter is entitled Safety Modification CN002/03/S which instructs the recalling firm's service engineers to visit each affected site and correct the software problem. A set of instructions were sent with the letter.
- Quantity in commerce
- 49 units
- Distribution
- The product was shipped to medical facilities in AL, CA, CO, FL, NE, IL, IN, MI, MN, NJ, NY, OH, PA, TN, and VA.
- Product codes and lots
- Part number 7502003